ANSI Measurement Guide
Oticon
Software: Oticon Genie, Oticon Genie2
Connection Device: miniRITE (FittingLINK)
Connection Device: miniRITE (FittingLINK)
- Detect wirelessly (FittingLINK)
- Connect HA
- Specify the receiver
- Select Settings
- Select Fitting Menu
- In Tools Menu, Select Technical Measurements
- Click OK for warning menu
- In Technical Measurements Window
- Select Full on Gain for ANSI
- Select ANSI Reference Gain for HA values
- Select User Settings for Directional and Distortion
- Oticon Genie (can select method of connection)
- Technical Data is present within software
Phonak
Software: iPFG, Phonak Target
Connection Device: iCube or NoahLink
Clinic HA:
Certena - IPFG
Exelia - IPFG
Naida V - Phonak Target
Solana - ?
Target
Connection Device: iCube or NoahLink
Clinic HA:
Certena - IPFG
Exelia - IPFG
Naida V - Phonak Target
Solana - ?
Target
- Open Phonak Target
- Set up with iCube (Wireless) or NoahLink (w/ cable)
- In Client & Session screen Click tools (bottom right)
- In Test settings tab, click Test settings under Hearing Instruments
- Select fitting device
- Click continue
- Select hearing instrument
- Select Gain Settings from this menu and run ANSI measurements
- For Real Ear Measurements, select Fitting, Fine Tuning
- Choose 3, 6, 10, or 16 for frequency selection
- Open Phonak IPFG
- Connect with NoahLink (w/cable)
- Click Detect
- Use settings from IPFG
- In Hearing System Tab, select measurement settings
- Save data for client to either the hearing instrument, database or both
- In measurement setting tab, select gain settings for ANSI measurements
- For Real Ear Measurements, select Fitting from the left menu and manual fine tuning
ReSound
Software: Aventa ReSound 2.95 (Old) ReSound Smart Fit/Aventa 3.12 (New)
Connection Device: Noahlink Wireless
Clinic HA: Linx 961
ANSI Step by Step
Connection Device: Noahlink Wireless
Clinic HA: Linx 961
ANSI Step by Step
- Open ReSound Smart Fit
- Click on Instrument tab
- Select Test Device
- Select Continue
- Select Continue
- On Test Device Window, select Full On Gain, Continue and Apply Test Settings
- If devices disconnect, select reconnect devices
- Place HAs in Verifit
- Select Test Box ANSI
- Select Start test for "Full on Gain" measurements
- Fit Patient for user settings
Starkey
Unitron
Software: U:Fit, Unitron TrueFit
Connection Device: iCube
ANSI Step by Step
Element 8
Connection Device: iCube
ANSI Step by Step
- Open Unitron TrueFit
- Connect HA via detect button
- Select Hearing Instrument, Test Mode
- Reconnect HA
- In Test Settings window, make ANSI test selections
Element 8
Widex
Software:
Connection Device:
Widex:
Bravo
Connection Device:
Widex:
Bravo