Evoked Potential

Cadwell Cascade Calibration

Calibration is a comparison of measurement values delivered by a device under test with those of a calibration standard of known accuracy, it correlates the readings of an instruments to those of a standard in order to check the instrument's accuracy. We continuously want to mark our instruments with a standard scale of readings. I recommend calibrating the equipment once a week to at least once a month.

  1. Never run amplifier diagnostics when connected to a patient
  2. Ensure that the latest software is downloaded and installed to your machine
  3. Connect your base to your laptop
  4. Connect your amplifer(s) and stimulation box to your base and turn on the base
  5. Connect an impedance test array to all four outoputs on your amplifer
  6. Open Cadwell
  7. Under Utilities click Amplifier Diagnostics
  8. In the first dropdown menu change from Gain to All
  9. In the second dropdown menu change from 1 to 3
  10. Click Begin
  11. The amplifier diagnostic will cycle through several waveforms
  12. A warning window will pop up to ensure that the patient is disconnected from stimulators, click OK
  13. The amplifier diagnostic will cycle through several waveforms again
  14. Once the amplifier diagnostic stops cycling, click on report
  15. When viewing the diagnostic report on the "Gain, Locut, Hicut, Noise" tab, make sure that all channels of your amplifer are showing the number 3. If any channel is showing a number less than 3, contact your system administrator for further amplifer diagnostics.
  16. Click Print Report, select the Diagnostics template, click Edit
  17. Add your name and the date to the report
  18. Save and upload the report
Evoked Potential
1640 Stockton Street ​#330071
San Francisco, California 94133
hello@evokedpotential.com